Signal cable connections GPro™ 500 TDL Series
Signal cable connections
Signal Description Cable no. Color
Power +24V Power 24 V, 5 W 1 Red
GND (Power) 2 Blue
RS485 A Interface M400 (RS485) 3 Green
RS485 B 4 Yellow
RS485 GND 5 Brown
4…20 mA pos Current input temperature 6 Purple
4…20 mA neg 7 Black
4…20 mA pos Current input pressure 8 Pink
4…20 mA neg 9 Grey
+24V Passive analog outputs 10 Red/blue
Out 1 11 Grey/pink
Out 2 12 White
TX+ Ethernet 13 White/yellow
TX– 14 Yellow/brown
RX+ 15 White/green
RX– 16 Brown/green
1
2 3 6 74 5
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Connections in the junction box
Connections to the GPro 500 –
Cable numbers below
Power from an external source or
optionally from the M400
RS485 from the M400
4…20 mA from temperature
sensor
4
20 mA from pressure sensor
Direct analog outputs (optional)
Ethernet