Mettler Toledo M400 Dokumentacja Strona 43

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 42
Transmitter Multiparameter M400 43
© 11 / 08 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M400
Printed in Switzerland 52 121 378
8.3 Analog Outputs
(PATH: Menu / Configure / Analog Outputs)
Enter configuration mode as described in Section 8.2.
Press the [ENTER] key to select this Menu, which lets you configure the 4 Analog Outputs.
Once analog outputs have been selected, use the and buttons to navigate between
configurable parameters. Once a parameter is selected, its setting can be selected per the
following table:
When an Alarm Value is selected, the analog output will go to this value if any alarm condition
occurs.
Parameter Selectable Values
Aout: 1, 2, 3 or 4 (default is 1)
Measurement: a, b, c, d or blank (none) (default is blank)
Alarm Value: 3.6 mA, 22.0 mA or Off (default is off)
The Aout type can be Normal, Bi-Linear, Auto-Range or Logarithmic. The range can be 4–20mA
or 0–20mA. Normal provides linear scaling between the minimum and maximum scaling l
imits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point
of the signal and allows two different linear segments between the minimum and maximum
scaling limits.
Enter the minimum and maximum Value of Aout.
If Auto-Range was selected then Aout max1 can be configured. Aout max1 is the maximum
value for the first range on Auto-Range. The maximum value for the second range on Auto-
Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt
for the number of decades as ”Aout1 # of Decades =2”.
The value for the Hold mode can be configured to hold the Last value or can be set to a Fixed
value.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.