
Transmitter M400 77
© 02 / 2010 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter M400
Printed in Switzerland 52 121 378
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.
8.4 Set points
(PATH: Menu / Configure / Set Points)
Enter Configuration mode as described in Section 8.1. “Enter Configuration Mode” and navigate
to the menu “Set Points” by using the or . key.
Press the [ENTER] key to select this menu.
Up to 6 setpoints can be configured on any of the measurements (a thru d). The possible Set-
point types are Off, High, Low, Outside and Between.
An ”Outside” setpoint will cause an alarm condition whenever the measurement goes above its
high limit or below its low limit. A ”Between” setpoint will cause an alarm condition to occur
whenever the measurement is between its high and low limits.
Enter the desired value(s) for the setpoint and press [ENTER]
Depending on the defined setpoint type, this screen provides the option to adjust the values for
the setpoint(s).
Press [ENTER] to proceed.
Out of Range
Once configured, the selected relay will be activated if a sensor Out of Range condition is detect-
ed on the assigned input channel. Select the setpoint and ”Yes” or ”No”. Select the desired relay
that will activate when the setpoint alarm condition is reached.
Press [ENTER]
Delay
Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously
for the specified length of time before activating the relay. If the condition disappears before the
delay period is over, the relay will not be activated.
Hysteresis
Enter the hysteresis as a percentage-value. A hysteresis value requires the measurement to re-
turn within the setpoint value by a specified percentage before the relay is deactivated.
For a high setpoint, the measurement must decrease more than the indicated percentage
below the setpoint value before the relay is deactivated. With a low setpoint, the measurement
must rise at least this percentage above the setpoint value before the relay is deactivated. For
example, with a high setpoint of 100, when this value is exceeded, the measurement must fall
below 90 before the relay is deactivated.
Press [ENTER]
Komentarze do niniejszej Instrukcji