4.3.2 TB2 Conductivity 4-E/2-E Analog Sensors
TB2 – Analog Sensors
Cond 4-E or 2-E
Terminal Function Color
A Cnd inner1* white
B Cnd outer1* white/blue
C Cnd outer1 –
D Not used –
E Cnd outer2 –
F Cnd inner2** blue
G Cnd outer2 (GND)** black
H Not used –
I RTD ret/GND bare shield
J RTD sense red
K RTD green
L Not used –
M Not used –
N Not used –
O Not used –
P Not used –
Q Not used –
* For third party Cond 2-E sensors may be jumper between A and B has to be installed.
** For third party Cond 2-E sensors may be jumper between F and G has to be installed.
4.3.3 Analog Sensors
TB2 – Analog Sensors
pH Redox (ORP)
Terminal Function Color* Function Color
A Glass transparent Platinum transparent
B Not used – – –
C No used – – –
D Not used – – –
E Reference red Reference red
F Reference** – Reference** –
G Solution GND** blue*** Solution GND** –
H Not used – – –
I RTD ret/GND white – –
J RTD sense – – –
K RTD green – –
L Not used – – –
M Shield (GND) green/yellow Shield (GND) green/yellow
N Not used – – –
O Not used – – –
P Not used – – –
Q Not used – – –
* Grey wire not used.
** Install jumper between F and G for ORP sensors and pH electrodes without SG.
*** Blue wire for electrode with SG.