7.5 Sensor temperature calibration
(only for analog sensors)
Enter Sensor calibration mode as described in section 7.1 “Enter Calibration Mode” and select
Temperature.
7.5.1 One-Point sensor temperature calibration
Select 1 point calibration. Slope or Offset can be selected with the 1 Point calibration. Select
Slope to recalculate the Slope factor M (Multiplier) or Offset to recalculate the offset calibration
factor A (Adder).
Enter the value for Point 1 and press [ENTER].
After a successful calibration, the calibration values are taken over (Adjust) or aborted
(Calibrate, Abort).
If “Adjust” has been chosen, the message “Calibration successful” is displayed. In any case yo
will get the message “Re-install sensor” and “Press ENTER” on the display. After pressing
“ ENTER” the M400 returns to the measuring mode.
7.5.2 Two-Point sensor temperature calibration
Select 2 Point as calibration type.
Enter the value for Point 1 and press [ENTER].
Enter the value for Point 2 and press [ENTER].