Mettler Toledo M300 ISM Dokumentacja Strona 61

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 60
Transmitter M300 61
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
imits and is the default setting. Bi-Linear will also prompt for a scaling value for the mid-point
of the signal and allows two different linear segments between the minimum and maximum
scaling limits.
Enter the minimum and maximum Value of Aout.
If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum
value for the first range on Auto-Range. The maximum value for the second range on Auto-
Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt
for the number of decades as ”Aout1 # of Decades =2”.
The value for the Hold mode can be configured to hold the Last value or can be set to a Fixed
value.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will
discard the entered values and return to the measurement display screen, selecting Yes will
save changes made.
Przeglądanie stron 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 121 122

Komentarze do niniejszej Instrukcji

Brak uwag