Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.
11.3.1.5 RgDiagnostic
Rg Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen
and select Channel A or B and Rg Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the
pH glass electrode measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the
pH glass electrode measuring input.
Pressing the [ENTER] key after entering Point 2 will bring up a confirmation screen. Select Yes
to save the calibration values and the successful Calibration is confirmed on the display. The
transmitter will return to the Measurement mode in approximately 5 seconds.
11.3.1.6 RrDiagnostics
Rr Diagnostic is performed as a two point calibration. Navigate to the Calibrate Meter screen
and select Channel A or B and Rr Diagnostic.
Enter the value for Point 1 of the calibration according to the resistor connected across the
pH reference measuring input. Press [ENTER] to begin the calibration process.
Enter the value for Point 2 of the calibration according to the resistor connected across the
pH reference measuring input.