Transmitter M300 75
© 12 / 11 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M300
Printed in Switzerland 52 121 389
10.2 PIDAuto/Manual
(PATH: MENU/PID Setup/PID A/M)
This menu allows selection of Automatic or Manual operation. Select Auto or Manual operation.
Pressing the [ENTER] key will bring up the Save Changes dialog.
10.3 Mode
(PATH: MENU/PID Setup/Mode)
This menu contains the selection of control modes using relays or analog outputs.
Press [ENTER].
10.3.1 PIDMode
This menu assigns a relay or analog output for PID control action as well as details of their op-
eration. Based on the control device being used, select one of the following three paragraphs for
use with solenoid valve, pulse input metering pump or analog control.
PulseLength – If using a solenoid valve, select ”Relays” and ”PL”, Pulse Length.
Choose the first relay position as #3 (recommended) and/or the second relay position
as #4 (recom mended) as well as the Pulse Length (PL) according to the table below. A longer
pulse length will reduce wear on the solenoid valve. The % ”on” time in the cycle is proportional
to the control output.
h
NOTE: All relays from #1 to #6 can be used for the controlling function.
1
st
RelayPosition
(#3)
2
nd
RelayPosition
(#4)
PulseLength(PL)
Conductivity
Controlling
concentrating reagent
feed
Controlling dilution water
Short (PL) provides
more uniform feed.
Suggested start point
= 30 sec
pH/ORP Feeding base Feeding acid
Reagent addition cycle:
short PL provides more
uniform addition of
reagent. Suggested start
point = 10 sec
Dissolved
Oxygen
Reverse control action
Direct acting control
action
Feed cycle time: short
PL provides more
uniform feed. Suggested
start point = 30 sec
Dissolved
Ozone
not recommended not recommended